Title: Calibration of the photon spectrometer PHOS of the ALICE experiment
Authors: Rath, Rutuparna
Roy, Ankhi
Sahoo, Raghunath
Keywords: Instruments;Electromagnetic calorimeter;Energy calibration;Gain calibration;High granularity;Mass resolution;Nonlinearity correction;Photon spectrometer;Time dependent;Calibration
Issue Date: 2019
Publisher: Institute of Physics Publishing
Citation: Acharya, S., Acosta, F. T., Adamová, D., Adhya, S. P., Adler, A., Adolfsson, J., . . . Zurlo, N. (2019). Calibration of the photon spectrometer PHOS of the ALICE experiment. Journal of Instrumentation, 14(5) doi:10.1088/1748-0221/14/05/P05025
Abstract: The procedure for the energy calibration of the high granularity electromagnetic calorimeter PHOS of the ALICE experiment is presented. The methods used to perform the relative gain calibration, to evaluate the geometrical alignment and the corresponding correction of the absolute energy scale, to obtain the nonlinearity correction coefficients and finally, to calculate the time-dependent calibration corrections, are discussed and illustrated by the PHOS performance in proton-proton (pp) collisions at s=13 TeV. After applying all corrections, the achieved mass resolutions for π0 and η mesons for pT > 1.7 GeV/c are σm π0 = 4.56 ± 0.03 MeV/c2 and σm η = 15.3 ± 1.0 MeV/c2, respectively. © 2019 CERN for the benefit of the Alice collaboration..
URI: https://doi.org/10.1088/1748-0221/14/05/P05025
https://dspace.iiti.ac.in/handle/123456789/8131
ISSN: 1748-0221
Type of Material: Journal Article
Appears in Collections:Department of Physics